| ¡Ü ÃøÁ¤Ç׸ñ |
|
|
| ÀÏ»êÈź¼Ò(CO) |
| ÀÌ»êÈź¼Ò(CO2) |
| ¿Âµµ(Temperature) |
| »ó´ë½Àµµ, À̽½Á¡, Àý´ë½Àµµ |
|
|
|
| ¡Ü Ư¡ |
|
|
| ¡¤ ÈÞ´ë¿ë Printer(Option)¿Í ¿¬°á »ç¿ë½Ã ÇöÀå¿¡¼ ¹Ù·Î µ¥ÀÌÅÍ Ãâ·Â °¡´É |
| ¡¤ CO, CO2, ¿Âµµ, ½Àµµ, À̽½Á¡, µ¥ÀÌÅÍ ÀúÀåÀÌ ÇÑ ´ë·Î °¡´É |
| ¡¤ ºñºÐ»ê Àû¿Ü¼± ¹æ½Ä(NDIR)À» ä¿ëÇÏ¿© ÃøÁ¤ ½Å·Úµµ°¡ ¿ì¼ö |
| ¡¤ ÃøÁ¤Ä¡¿¡ ´ëÇÑ Åë°èó¸®(Æò±Õ, ÃÖ´ëÄ¡, ÃÖ¼ÒÄ¡) ±â´ÉžÀç |
| ¡¤ ¹Ì±¹ÀÇ NIST ÀÎÁõ¼¸¦ ȹµæÇÑ ½Å·Úµµ°¡ ÀÔÁõµÈ Á¦Ç°ÀÓ |
|
|
|
| ¡Ü »ç¾ç |
|
|
|
|
|
|
Model |
TSI-8762 |
|
CO2 |
|
Sensor Type |
Dual-wavelength NDIR (non-dispersive infrared) |
|
Range |
0 to 5000 ppm |
|
Accuracy |
¡¾3.0% of reading or ¡¾50 ppm, whichever is greater |
|
Resolution |
1 ppm |
|
Temperature (Models 8760 and 8762) |
|
Sensor Type |
Therrmistor |
|
Range |
32 to 140¡ÆF (0 to 60¡ÆC) |
|
Accuracy |
¡¾1.0¡ÆF (¡¾0.6¡ÆC) |
|
Resolution |
0.1¡ÆF (0.1¡ÆC) |
|
Optional Secondary Temperature Probe (Models 8760 and 8762) |
|
Sensor Type |
RTD |
|
Range |
-40 to 248¡ÆF (-40 to 120¡ÆC) |
|
Accuracy |
¡¾1.0¡ÆF (¡¾0.6¡ÆC) |
|
Resolution |
0.1¡ÆF (0.1¡ÆC) |
|
Relative Humidity (Models 8760 and 8762) |
|
Sensor Type |
Thin-film capacitive |
|
Range |
5% to 95% RH |
|
Accuracy |
¡¾2.0% RH |
|
Resolution |
0.1% RH |
|
Dew Point (Models 8760 and 8762) |
|
Range |
5 to 120¡ÆF (-15 to 49¡ÆC) |
|
Resolution |
0.1¡ÆF (0.1¡ÆC) |
|
Wet Bulb Temperature (Models 8760 and 8762) |
|
Range |
40 to 140¡ÆF (5 to 60¡ÆC) |
|
Resolution |
0.1¡ÆF (0.1¡ÆC) |
|
Absolute Humidity (Models 8760 and 8762) |
|
Range |
0.0000151 to 0.00491 lbs moisture/ft3 air |
|
(0.24 to 79.0 g moisture/m3 air) |
|
Humidity Ratio (Models 8760 and 8762) |
|
Range |
0.00014 to 0.187 lbs moisture/lb dry air |
|
(0.14 to 187 g moisture/kg dry air) |
|
% Outside Air (Models 8760 and 8762) |
|
Range |
0 to 100% |
|
Resolution |
0.1% |
|
CO (Model 8762) |
|
Sensor Type |
Electro-chemical |
|
Range |
0-500 ppm |
|
Accuracy |
¡¾3.0% of reading or ¡¾3 ppm, whichever is greater |
|
Resolution |
0.1 ppm |
|
Response Time |
¡¡ |
|
Operating Temperature |
|
Models 8732 and 8760 |
41 to 158¡ÆF (5 to 70¡ÆC) |
|
Model 8762 |
41 to 113¡ÆF (5 to 45¡ÆC) |
|
Storage Temperature: -4 to 158¡ÆF (-20 to 70¡ÆC) |
|
Logging Capability (Models 8760 and 8762) |
|
Range |
Up to 14,000 samples and 1,000 test ID¡¯s |
|
Intervals |
5 sec, 10 sec, 15 sec, 20 sec, 30 sec, 60 sec, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min |
|
Probe Dimensions (Model 8760) |
|
Length |
5.71 in (14.5 cm.) |
|
Diameter |
0.75 in (1.9 cm.) |
|
Probe Dimensions (Model 8762) |
| Length |
6.50 in. (16.5 cm) |
| Diameter |
0.75 in. (1.9 cm) |
| Weight (with batteries) |
| Model 8732 |
0.76 lbs (0.34 kg) |
| Models 8760 |
1.16 lbs (0.53 kg) |
| Model 8762 |
1.16 lbs (0.53 kg) |
| Power: Four AA-size batteries (included) or AC adapter (Included with Models 8760 and 8762) |
| Printer Interface |
1200 baud rate, serial connection |