SP-2000 Visible Spectrophotometer

SPectrum SP-2000 Spectrophotometer is compact, easy to use and features include a 6nm bandwidth, wavelength range of 325-1000nm,T%
/Abs/Conc. test mode, automatic zeroing and blanking, a four-position cell holder, a free of charge

Windows®-based T%/A/C Application software for easy data collection and export to Excel?, alignment-free lamp change (no tools required), optional external serial printing, a large sample compartment of instrument for various accessories. The print control is built-in to the instrument. Just push the PRINT button on the front panel to send data to your printer. Or you can interface with your Windows 9X or Windows 2000 computer to create a system for test data manipulation, management and record keeping. Optional accessories include a 10 mm single-cell or round test tube holder, a round test tube holder for 8-22 mm diameter test tubes, a water-jacketed flow-cell holder and panel and a long path cell holder for up to 100 mm long cell.
   The SP-2000 Visible spectrophotometer is ideal for various applications such as school, college, clinic, water, wastewater lab and other quality control fields.


Features:
Auto 100%T/0A and 0%T.
A Large digital LCD displays test results of %T, Abs or Cone.
Sample compartment contains a four-position cell holder.
Silicon photodiode detector covers the entire wavelength, eliminating the need to change detector.
6nm spectral bandwidth provides the sensitivity and performance required for almost all applications.
Built-in stray light filters.
Built-in RS-232C interface
Able to test micro samples.

Specifications:
Optical System: Littrow Mounting, single beam, 1200l/mm grating optical system,
Wavelength range: 325-1000nm
Spectral bandwidth: 6nm
Wavelength accuracy: ÷2nm
Wavelength readability: ÷1nm
Stray radiant energy: Less than 1% at 340nm
Photometric range: 0-100.0%T, 0-1.999A, 0-1999C(0-1999F)
Photometric accuracy: ÷1%T
Data Output: Analog, RS-232C
Light Source: 6V/10W Tungsten Halogen Lamp
Sample Container: four-position cell holder,
Detector: High sensitivity silicon photodiode
Standard Accessories:
Instruction manual, Dust cover, 10x10x45 rectangular cell, %Trans/Abs/Conc Application software for Win 9x

 